Starting up a HYDROCELL column:

Whether from storage or fresh from the manufacturer, the baseline can 
drift when you first use a Hydrocell Ion Exchange column. The column needs to be activated. The procedure 
of activating a 4.6 mm Hydrocell column is as follows:

1.      Flush with 100% water to remove any salts from your HPLC system,
then 80% methanol in water at a flow rate of 3 mL / min. for 10 
minutes (Do not repeat this procedure) .

 

2.       Flush the HPLC system with 100% water, and then flush the 
column with 20% methanol in 1 M sodium chloride solution at flow rate 
of 3 mL / min for 10 minutes.

 

3.      Flush with your B Buffer of choice at flow rate of 3 mL / min for 10 minutes.

 

4.      Equilibrate with your starting buffer of choice at a flow rate of 2 mL / min.

 

5.      Run a 10 minute blank gradient from 0 -100 % B buffer at flow rate of 2 mL / min.

 

*If the column baseline is still not straight or has spike peaks, please repeat the procedure from

2 to 5